Specifications
Measurement System: Discrete, random access, multi-test analysis Number of tests on board: 30 items max. Throughput: 120 tests/hour Analytical method: Endpiont, Rate assay, ISE (Option) Calibration curve: Factor, Linear, Spline, Logit-log1, Logit-log2
Sample Handling Contents of test: Clinical chemistry, Immuno-assay Sample container: Cup, Speed Tube II Sample number: For patient sample: 6 positions, For QC sample: 3 positions Sample volume: 2-25 m l Automatic dilution: Dilution ratio: 50 max. Sample information: Position and barcode
Reagent Handling Reagent tray: 30 sectors Reagent volume: 20-200 m l Reagent information: 2-dimensional barcode
Reaction Cuvette material: Plastics (semi-disposable) Reaction liquid volume: 100-200 m l Reaction temperature: 37ºC ± 0.1ºC Wavelength: 340-800 nm Light source: Tungsten halogen lamp Cuvette washing: Automatic washing system Water supply: Max 1.0 L/hour
User interface Touch screen: Color screen Quality control: Within day, day-to-day validation Data storage capacity: 6 months System interface: LAN (TCP/IP)
Installation condition Electrical requirements: AC 85-265 V (47-63 Hz) (Power Consumption: 300 VA) Physical temperature: 15 ºC-30 ºC Dimensions: 520 (W) x 605 (D) x 435 (H) Weight: Approx. 32 kg
Option ISE module Patient report system Vacuum blood collection tube: Speed tube II Water purifying system External printer
|